Pro-MAM Performance - Pro Project - 03-601
The phantom designed to be used for routine quick and ongoing checks of imaging performance of mammography systems. It allows to evaluate particularly those aspects which are most liable to deterioration.
Technical data (can be modified to customer specifications):- 240mm diameter semicircle
- first half of the test module contains:
- 6 groups of 10mm long multi-directional filaments of the following diameters: 0.40, 0.35, 0.30, 0.25, 0.225, 0.20 mm
- 6 groups of microcalcifications in the following ranges: 354-224, 283-180, 226-150, 177-106, 141-90, 106-63 μm
- 6 groups of 3 low contrast details, 3mm in diameter, producing nominal contrast at 28kVp of: 0.04, 0.03, 0.02, 0.015, 0.01, 0.005 (each contrast is presented 3 times)
- second half of the test module contains:
- clinically realistic breast tissue (not tissue equivalent to but simulating the appearance of the breast tissue) containing 6 groups of low contrast detailssame specifications as in the other half
- 6 PMMA plates 10mm thick
- 2 PMMA plates 5mm thick
- carrying case
- Complies with:
- IEC 61223-3-2
- CE certified
- the Manual provides detailed guidelines for carrying out each test, results assessment and registration