Extended Oval PET Phantom™
			Simulates the distribution of radioactivity throughout the long body torso*
			
			The Extended Oval PET Phantom Set™ is designed to simulate the distribution of radioactivity throughout the long body torso*. It consists of a set of two 40 cm long cylinder sections with oval shape cross section to simulate the long body torso. A single 40 cm section fills the axial FOV of a large camera
Main Applications:- Evaluation of cancer imaging using camera-based and dedicated PET systems
- Measurements of contrast, signal-to-noise ratio using hot inserts
- Evaluation of the effects of out-of field-of-view activiites
- Evaluation of attenuation and scatter correction
- Research
| Specifications: | |
| Outside lateral dimensional of cylinder | 36 cm | 
| Inside lateral dimension of cylinder | 34 cm | 
| Outside anterior-posterior dimension of cylinder | 21 cm | 
| Inside anterior-posterior dimension of cylinder | 19cm | 
| Wall thickness | 1 cm | 
| Height of each cylinder section | 40 cm | 
| Total cylinder height | 80 cm | 
Model PET/EX-OV/P
* Turkington TG, Williams NE, Hamblen SM and Coleman RE. Regional FDG uptake, attenuation, and geometry measurements for whole body phantom design. J Nucl Med 40:281P, 1999.
 
		 
			